ET II General - Set 1

Rating
0 0

There are no comments for now.

to be the first to leave a comment.

2. In 1824 Arago showed that a rotating copper disc exerts a force on a nearby magnet. This was early evidence of:
3. Which pioneer used an induction balance in 1879 to tell different metals apart, an early form of eddy current sorting?
5. Using δ(mm) = 50·√(ρ/(f·μr)), what is the standard depth in 304 stainless steel (ρ ≈ 72 µΩ·cm, μr = 1) at 100 kHz?
6. Eddy current density at a depth of two standard depths is approximately what percentage of the surface density?
7. If the test frequency is raised by a factor of 100 with everything else unchanged, the standard depth:
8. At a depth of two standard depths, eddy currents lag the surface currents in phase by approximately:
9. A metal has a resistivity of 3.45 µΩ·cm. Its conductivity is closest to:
10. Which of the following materials is diamagnetic, with a relative permeability slightly below 1?
13. When a surface coil is placed on a non-magnetic conductor, the coil's resistive component typically ___ and its inductive reactance typically ___.
14. A coil has an inductance of 80 µH. What is its inductive reactance at 100 kHz?
15. A coil has R = 40 Ω and XL = 40 Ω. Its phase angle θ = arctan(XL/R) is:
16. An AC waveform has a period of 10 µs. What is its frequency?
18. On a normalized impedance plane, an unsaturated ferromagnetic steel part usually places the operating point:
20. Using fg = 5066/(σ·μr·d²), a brass rod with σ = 16 MS/m and d = 1.0 cm has fg ≈ 317 Hz. Which test frequency gives f/fg = 10?
21. If the diameter of a bar doubles while its material stays the same, its characteristic frequency fg:
22. Compared with crack detection in bar, conductivity sorting with an encircling coil generally uses:
23. A 15.0 mm diameter bobbin probe is used in a tube with an inside diameter of 17.0 mm. What is the fill factor?
26. A crack signal measures 1.2 V peak-to-peak and the background noise 0.3 V. The signal-to-noise ratio is:
27. An operator doubles the instrument gain to make a weak flaw signal easier to see. The signal-to-noise ratio:
28. Scanning a weld near a running welding set, the display shows fast, spiky hash superimposed on every signal. Which filter setting is aimed at this?
29. An increase in instrument gain of 20 dB multiplies the signal amplitude by about:
31. A differential bobbin probe passing a small hole typically produces what pattern on the impedance-plane display?
35. After a steel part has been tested with DC magnetic saturation, what may be needed before it moves on to later processes?